Low Pressure In Diffusion Call ASP
Technical Service
Problem/Condition
The system did not reach atmospheric pressure by the end of the
diffusion stage.
or
For the second consecutive cycle, diffusion pressure is under 6 Torr.
There is insufficient H2O2 for diffusion or it is being absorbed by the
load.
Action
F
Check the vacuum switch
F
Check the vent valve replace if necessary.
F
Run an empty chamber cycle.
F
Check the door and chamber temperatures.
F
Test the injector valve function.
F
Test the pneumatics system.
Low pressure in diffusion. Repackage
load. Restart sterilizer.
Problem/Condition
Diffusion pressure is under 6 Torr. There is insufficient H2O2 for
diffusion or it is being absorbed by the load.
Action
F
Check for absorbent materials in the load.
F
Repackage the load.
F
Check the cassette.
F
Restart the sterilizer.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-11