Replace the Control Enclosure
1. Attach the control enclosure to the frame.
2. Plug in the control harnesses from the subsystems as labeled on the
enclosure.
3. Install the circuit boards removed from the old enclosure into the
new enclosure.
4. Plug in the AC power.
Remove the Power Supplies
The Block 2 procedures are slightly different and are included in the
steps below.
1. Open the control enclosure side panel.
2. Remove the 4 nuts fastening the 12.85 V power supply On Block 2
remove the 4 screws mounting the power supply assembly.
3. Remove the standoffs from the 5 V power supply. (Not present on
Block 2.)
4. Remove wires from terminal block. (Block 2. Unfasten the power
supply connections.)
Replace the Power Supplies
1. Fasten the wires to the terminal block. (Block 2. Reconnect the
power supply).·
CAUTION Be sure you connect the wires correctly. See the
block 1.8 wiring diagram on the following page for
information.
2. Reattach the power supplies using the standoffs (Block 1.8, 5 V), or
the 4 nuts (Block 1.8 12 V) or the 4 screws (Block 2)
3. Replace the control enclosure side panel.
Control Subsystem Replacement
User Interface and Control Subsystems
ASP Service Guide rev D.
STERRAD® 100 Sterilizer
6-25
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......