11. Route the cables over the pulleys. It is important that the cables fit
into the correct tracks on the pulleys: the left track of the front
pulley, and the right track of the rear pulley.
12. Adjust the door cables so that there is 15 - 20 mm clearance at the
air cylinder hex nut when the door is in the locked position. Adjust
the upper door stops to extend 21.5 mm.
13. Verify that there is a gap between the door and the rail. Install and
reconnect the door over-center flag (the small flag packed loose with
the door). Install the door sensor harness, thermistor, ground, heater
harnesses.
14. After adjustments are made, increase the air pressure to 60-80 PSI
and verify door closing speed and door operation. When the door
speed and operation is satisfactory, lock the adjusting rings on the
speed control valves. Lock all jam nuts to the cable ends.
15. Remove and replace the door seal o-ring.
16. Reinstall all panels.
17. If you are installing a show door, replace the front panel with the
“show panel” included in the kit.
18. Perform a leak back test.
Replace the Door with a Show Door
Use the same procedures to install the show door as use with a standard
door. The only difference is in the handling of the harnesses and cables.
1. Install the door sensor harness only.
2. Tywrap the thermistor, ground and heater harnesses so that they
don’t interfere with the movement of the door. They are not used on
a show door.
Chamber Heaters
Removal/Replacement
Remove the Heaters
HAZARDOUS CHEMICALS!· The primer and adhesive used in
this procedure are caustic and have hazardous
vapors. Be sure you are using them in a well
ventilated room and that you protect your clothes
and skin. You MUST wear side shield safety glasses,
rubber gloves and a protective smock when using
RTV adhesives and primers
Following are the steps for removing a defective heater, cleaning the
chamber of adhesive residue and installing a new heater. In addition,
here is a list of additional supplies you will need to remove and install
the chamber heaters.
Door and Chamber Subsystem
Removal/Replacement
7-18
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......