8. Adjust and tension the door cables so that door hangs evenly and
operates smoothly.
9. Adjust the cables to give 15 to 20 mm clearance at the air cylinder
hex nut when the door is in the locked position. See the diagram.
10. Adjust both upper door stops to extend 21.5 mm (start position).
11. Adjust the door stops so that the door over center mechanism
extends and closes the door against the O-ring evenly. View the air
gap between the door and the chamber face from top and bottom to
confirm correct alignment.
4
4
Note:
The door links should be parallel to each other, and at
horizontal or pointing slightly up at the rear.
12. Reconnect the air lines to the door cylinder. Replace the front panel.
Perform the alignment procedures for Door Top, Bottom, And Over
Center Sensors.
Figure 7-4. Align Left Guide Rail to Frame
Calibration
Door and Chamber Subsystem
ASP Service Guide rev D