M
750 mTorr
Max. pressure rise in 10 min.
(38-48 min.)
C
525 mTorr
Plasma high pressure alarm
C
475 mTorr
Plasma low pressure alarm
C
690 Torr
Vent (atmospheric pressure switch)
C
1 Torr
Vent pump-down threshold
RF Delivered Power
Directional coupler in RF amplifier board.
Control/
Monitor
Alarm Limits
Process Equipment
Specification
M
2.92 Volts
Minimum allowable power (365 W)
Temperature
Thermistors are located on the chamber and door to monitor
temperature.
Control/
Monitor
Alarm Limits
Process Equipment
Specification
C
40°C
Lower limit
C
50°C
Upper limit
M
60°C
Temperature alarm threshold
Software Overview
Software controls the STERRAD® 100 Sterilizer to ensure the most
efficient sterilization process as well as ease of use and operator safety.
The software manages the sterilization process throughout the five
stages: vacuum, injection, diffusion, plasma, and vent. The software also
monitors pressure, temperature, and other factors to determine when to
move from one stage to another, what to adjust in the various stages,
and when to abort the process if there are problems. The software
controls the machine in the monitor state including the processing of
hydrogen peroxide cassettes.
Theory of Operation-Overview
System Operation
ASP Service Guide rev. D
STERRAD® 100 Sterilizer
2-13
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......