Plasma not lit. Call ASP Technical
Service.
Problem/Condition
RF delivered power is less than 200 W but greater than 100 W.
Action
F
Check shelves and make sure they are not in contact with the
electrode or chamber door.
F
Check the RF system.
Please Inspect Load Carefully See Op
Manual Appendix A
Problem/Condition
Hydrogen peroxide may be present on wet items.
Action
F
HANDLE THE LOAD WITH GLOVES
as H
2
O
2
may be present.
F
Dry load.
F
Repackage the load.
F
Restart the sterilizer.
PLEASE REMOVE CASSETTE
There is no printed message for this condition.
Problem/Condition
A cassette is detected by the reflective sensor, but is not under the roller.
Action
F
Remove the cassette and reinsert.
F
If the message repeats, insert a new cassette.
F
Check the injector assembly connections.
System Messages
Using System Messages
3-22
STERRAD® 100 Sterilizer
ASP Service Guide rev D