PROCESS COMPLETE CALL TECHNICAL
SERVICE
F
Level 1 PM Interval Past Due. Call ASP Technical Service.
F
Level 2 PM Interval Past Due. Call ASP Technical Service.
Problem/Condition
F
The Level 1 PM is past due.
or
F
The Level 2 PM is past due.
Action
F
Perform a Level 1 PM
F
Perform a Level 2 PM
Process Time Exceeded. Call ASP
Technical Service
Problem/Condition
F
The entire sterilization process exceeded the maximum time
parameters.
F
The system locks up in the vacuum or injection stages. Alarm light is
activated.
Action
F
Check the injector pump optics.
F
Check the throttle valve connection
F
Test the throttle valve.
F
Check the power supplies (5.2 V)
F
Replace the microprocessor board.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-29