Step
Cycle Parameters
Short Cycle Time
Long Cycle Time
1
Vacuum to 700 mTorr
5 minutes (approx)
5 minutes (approx)
2
Pre-Plasma at 800 mTorr
10 minutes
15 minutes
3
Vent to 760 Torr
1 minute
1 minute
4
Vacuum to 400 mTorr
5 minutes (approx)
5 minutes (approx)
5
Injection 1 at 400 mTorr
6 minutes
6 minutes
6
Diffusion 1 at 760 Torr
2 minutes
10 minutes
7
Vacuum to 500 mTorr
5 minutes (approx)
5 minutes (approx)
8
Plasma at 500 mTorr
2 minutes
2 minutes
9
Vacuum to 400 mTorr
0.1 minute
0.1 minute
10
Injection 2 at 400 mTorr
6 minutes
6 minutes
11
Diffusion 2 at 760 Torr
2 minutes
10 minutes
12
Vacuum to 500 mTorr
5 minutes (approx)
5 minutes (approx)
13
Plasma at 500 mTorr
2 minutes
2 minutes
14
Vent to 760 Torr
0.3 minute
0.3 minute
TOTAL TIME (approximate) 51 minutes
72 minutes
STERRAD 100 S Theory of Operation
System Operation
ASP Service Guide rev. D
STERRAD® 100 Sterilizer
2-23
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......