No cassette in sterilizer. Repackage
load. Restart sterilizer.
Problem/Condition
F
The sterilization process canceled because there is no cassette in the
sterilizer.
F
Cycle was started or restarted without a cassette.
F
Injection was interrupted at the 10th cell of the cassette.
Action
F
Insert a new cassette.
F
Repackage the load.
F
Restart the sterilizer.
NO DISPLAY
Software anomaly SP = XXXXX Call ASP Technical Service
Problem/Condition
There is a software error. The printed message contains the memory
address of the error.
Action
F
Test the controller.
F
Check the power supplies .
F
Replace the software.
System Messages
Using System Messages
3-16
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100