10. Verify that the software parameters are displayed on the LCD. These
parameters include the month and year, the words STERRAD 100
and the software part number and CS number. The parameters
change with each software release.
11. Remove the firmware if the parameters are not verified. Discard the
incorrect EPROM and repeat with a new EPROM
12. Enter the EDIT mode and set the correct time, date, serial number,
and total machine cycles as previously shown in the calibration
section.
Control Subsystem Replacement
User Interface and Control Subsystems
ASP Service Guide rev D.
STERRAD® 100 Sterilizer
6-27