Over pressure in injection. Repackage
load. Restart sterilizer.
Problem/Condition
Chamber pressure is above 12 Torr in the injection stage.
Action
F
Inspect the load for plastic items that may be out-gassing.
F
Replace the cassette.
F
Inspect system for leaks
F
Repackage the load.
F
Restart the sterilizer.
Over Pressure In Injection See Op
Manual Appendix A
Problem/Condition
The chamber pressure is above 14 Torr in the injection stage of the
STERRAD 100 S System.
Action
F
Replace cassette.
F
Inspect injection system.
F
Perform leak back test.
F
If leak back test fails check for leaks.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-21