F
Check that door is aligned horizontally.
Corrective Action
F
Make adjustments to stops and guide cables as necessary.
F
Replace door solenoid
F
Replace pneumatics board.
Display
OPEN CHAMBER DOOR
W/ START OR NEXT W/ CANCEL
Action
F
Press
START
.
F
Verify that chamber door opens.
F
Adjust speed control valve on door cylinder.
F
Adjust and secure speed control valves on pneumatic control
assembly, so that door opens smoothly in approximately 5 seconds.
Corrective Action
F
Make adjustments to stops and guide cables as necessary.
F
Replace door solenoid
F
Replace pneumatics board.
Display
CLOSE CHAMBER DOOR
W/ START OR NEXT W/ CANCEL
Action
Press
CANCEL.
AutoTest
Injection and Pneumatics Tests
B-20
STERRAD® 100 Sterilizer
ASP Service Guide rev D