Action
Provides instructions for completing the tests and certifications required.
Text in
Bold
indicates the switch to press to perform the test or continue
to the next test. Pressing
START
,
CANCEL
or
LONG CANCEL usually
advances to the next text or display. Press
LONG CANCEL
to skip or exit
out of a test or menu.
Corrective Action
This information is a list of things to do if the system is not working
correctly.
F
Perform corrective action in the sequence given until failure is found
then restart test sequence given.
F
Do not replace items if they are not the cause of failure.
F
Repeat any test that shows an anomaly, or where you adjust or
replace any component or part.
User Interface Tests
Initial Start-Up Display
AUTOTEST Block 1.8/2.0
Next menu w/ long cancel
0239810 Rev A, 7/19/93
Next menu w/ long cancel
Action
F
Verify that software part number and revision are correct.
F
Verify display is working and has proper contrast.
F
Adjust contrast with R1 pot on rear of display board.
F
Press a
LONG CANCEL
.
User Interface Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-3
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......