Action
Press
CANCEL
for the next sensor test.
Corrective Action
F
Check the connections in the injector pump.
F
Check the floating guides.
F
Adjust the cassette optics.
F
Replace injector pump.
Display
Normal display
TURN CHUTE SWITCH ON
NEXT SENSOR W/ CANCEL
or with chute switch pressed.
CHUTE SWITCH ON
NEXT SENSOR W/ CANCEL
Action
F
Manually press down the chute switch. Beeps sound when the
chute switch is closed. No beeps sound when the switch is opened.
F
Press
CANCEL
for the next sensor test.
Corrective Action
F
If no beeps, check the wiring and connections.
F
Replace switch.
F
Replace the pneumatic board.
F
Replace the digital I/O board.
Display with collection box removed.
TURN BOX SWITCH ON
NEXT SENSOR W/ CANCEL
or with collection box in place.
BOX SWITCH ON
NEXT SENSOR W/ CANCEL
Sensor Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-11