Door Sensor Tests
Display
The door tests are normally run with the door open. Alternate displays
are for tests with the door closed.
BLOCK DOOR TOP SENSOR
NEXT SENSOR W/ CANCEL
or with the door closed.
DOOR TOP SENSOR BLOCKED
NEXT SENSOR W/ CANCEL
Action
F
If the door top sensor is not blocked, use an object to block the door
top sensor, or manually close the door. A beep occurs when the
sensor is blocked.
F
Press
CANCEL
for the next sensor test.
Corrective Action
F
If no beep, check the wiring and connections.
F
Replace the sensor board.
Display
DOOR BOTTOM BLOCKED
NEXT SENSOR W/ CANCEL
or with the door closed.
BLOCK DOOR BOTTOM
NEXT SENSOR W/ CANCEL
Action
F
If the door bottom sensor is not blocked, use an object to block the
door bottom sensor. A beep occurs when the sensor is blocked.
F
Since the door is usually open at this time, the sensor is blocked by
the flag. Use the blue override button on the pneumatic solenoid to
verify the sensor is working, by moving the door up.
F
Press
CANCEL
for the next sensor test.
Sensor Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-9
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......