TEMP SENSOR DISCONNECTED NO MORE
CYCLES CAN RUN
Temperature Sensor Disconnected. Call ASP Technical Service
Problem/Condition
The temperature sensor is disconnected or has failed.
Action
F
Check the temperature sensor and make sure it is connected and
operating normally.
F
Check the door heater and thermistor, replace if necessary.
Temperature has not risen.
Problem/Condition
The system has been below temperature for more than 90 minutes.
Action
F
Close the door and allow the unit to warm up for at least an hour.
F
Run an empty chamber cycle.
F
Check door heater and thermistor. If recorded temperature is greater
than 40° C, replace the thermistor.
Temperature has not risen. Call ASP
Technical Service.
Problem/Condition
The temperature of the door is below 40° C and has remained so for
more than 1 hour.
Action
F
Close the door and allow the unit to warm up for at least an hour.
F
Run an empty chamber cycle.
F
Check the door heater and thermistor. If recorded temperature is
greater than 40° C, replace the thermistor.
System Messages
Using System Messages
3-34
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......