CLOSE DOOR TIME-OUT CALL TECHNICAL
SERVICE
There is no printed message for this condition.
Problem/Condition
More than 15 seconds passed before the door top sensor was activated.
Action
F
Check the door top sensor and speed control valve
F
Verify pneumatic pressure. Observe gauge (60 to 80 PSI)
Control Panel is Blank
Problem/Condition
5 VDC power supply checks low at 4.68 VDC; analog VDC is low at
approximately 1.880. Equipment running fine otherwise. During
adjustment of 5 VDC power supply equipment shuts down at 4.80 VDC.
Action
F
Check contrast.
F
Attempt to enter and then exit the Edit mode.
F
Check voltages.
F
Perform a Master Reset.
F
Start cycle, cancel it and recheck the voltages.
F
Replace the 5 V power supply.
F
Replace the control enclosure.
CYCLE 1 SHORT CYCLE
YES(START)—NO(CANCEL)
F
If you are not processing a flexible scope, press
START
to start a
short cycle.
F
If you are processing a flexible scope, press
CANCEL
to go to the
CYCLE 2
(long cycle) option.
System Messages
Using System Messages
3-4
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......