Action
Press a
LONG CLOSE.
Display
INPANEL W/ START or
NEXT Menu W/ LONG CANCEL
Action
To repeat the test, press
START
, otherwise press a
LONG
CANCEL
for
the next menu.
Corrective Action
F
Check connections.
F
If any switch is not operating correctly, the entire membrane panel
must be replaced.
Lamps Test
This test steps through checking the various display panel lights. The
corrective action is the same for any light malfunction and is presented
at the end of the test.
Display
OUTPANEL W/ START. or
NEXT Menu W/ CANCEL
Action
Press
START.
Display
ON/OFF LIGHTS W/ START
NEXT LIGHT W/ CANCEL
Action
F
Press
START.
F
Verify both alarm lamps are on.
F
Press
START
to turn light off.
F
Press
CANCEL.
AutoTest
User Interface Tests
B-6
STERRAD® 100 Sterilizer
ASP Service Guide rev D