F
Remove connector from capacitance manometer (Baratron) pressure
sensor and attach the adjustment tool to the connector.
F
Adjust trimpot on tool so that value in LED is approximately 4.000 V.
(or similar number that is easily divided by 2.)·
F
Press
CANCEL.
Corrective Action
F
If no reading, or reading is not steady, check wiring for noise or bad
connection.
4
4
Note
: Channels 1-6 are not tested at this time. Press CANCEL
to skip.
Display
CHAN #1 RF REFLECT POWER
CONTINUE W/ START
Display
0.587 V., REFLECT POWER
NEXT INPUT W/ CANCEL
Display
CHAN #2 RF DEL. POWER
CONTINUE W/ START
Display
0.589 V., DEL. POWER
NEXT INPUT W/ CANCEL
Display
CHAN #3 FLOATING INPUT
CONTINUE W/ START
Action
Press
START.
Display
4.095V., FLOATING POINT
NEXT INPUT W/ CANCEL
AutoTest
Analog Test
B-30