Door Top, Bottom, And Over Center
Sensors
1. Remove power to system. Remove the system panels. Disconnect and
mark the air lines to the door cylinder so it can be manually operated.
2. Manually move the door to the top (not the over center) position and
verify that the door flag just blocks the optical path of top sensor.
3. Adjust the flag (mounted on door) as necessary to achieve vertical
positioning within the door sensor PCB optical path.
4. Adjust the sensor PCB bracket (mounted on the guide rail) as
necessary to ensure the flag passes well into the light path, but does
not hit the sensor PCB.
5. Verify that the flag passes midway between the optical devices located
on the sensor PCB. Adjust sensor PCB as necessary.
6. Manually move the door to the bottom (not the service stop) position
and verify the door flag blocks the optical path of the bottom sensor.
7. Repeat previous adjustment steps for bottom sensor alignment.
4
4
Note
: There is only one door flag used for both top and bottom
sensors. Adjusting this device for bottom position affects
the previously set top position. Recheck top sensor if the
flag must be moved.
8. Move the over center flag up until the optical path is just blocked,
then lower the flag approximately 2.0 mm.
9. Verify that the flag passes midway between the optical devices.
Adjust sensor PCB as necessary.
10. With no pressure applied to the door over center mechanism, verify
the door over center flag does not block the sensor.
11. Reconnect air lines to door cylinder and check door operation with
normal firmware. Ensure that the door over center mechanism is
functioning properly by blocking door and observing the reversing
action.
4
4
Note
: The door over center mechanism should actuate with
approximately five pounds (2.5 kilos) of pressure applied
to the top of the door.
Door Open Close Speed
1. Remove the system panels. Verify that the air pressure is set
correctly to cut-in at 60 PSI, and to cut-out at 80 PSI.
2. Loosen the speed control valve locking nut on the control valve
located on the exhaust (lower) side of the door cylinder.
Door and Chamber Subsystem
Calibration
7-6
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......