Power interrupted. Repackage load.
Restart sterilizer.
Problem/Condition
F
The power was interrupted, even momentarily, during injection,
during the first 4 minutes in diffusion or plasma stages; or for more
than 1 minute during diffusion.
F
If power is out for more than one minute during the vacuum stage,
the cycle restarts.
Action
F
Repackage the load.
F
Restart the sterilizer.
Power Interrupted See Op Manual
Appendix A
Problem/Condition
F
Power was interrupted for more than one minute during and
pump-down or either diffusion stage.
or
F
Power was interrupted for any length of time during application of
plasma or during the injection stage.
Action
F
Repackage the load and restart the sterilizer.
If the problem persists:
F
Check incoming power supply. Verify that phase is correct.
F
Check for loose connections.
F
Replace enclosure.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-25