Align Chamber to Right Guide Rail
1. Remove the screw and cable from the right upper door stop bracket
to give clearance for the gage.
2. Loosen, do not remove, the three mounting bolts securing the
chamber to the right and left guide rail.
3. Using T.2266 align the chamber with the right guide rail. See the
diagram. Adjust the chamber mounting eyebolt located at the rear of
the chamber. This sets the chamber face parallel to the right guide
rail, giving a slip fit.
4. Tighten the three bolts securing the right guide rail to the chamber.
5. Tighten the stand-off between the guide rail and the frame.
6. Using gage T.2266 verify that the alignment of the chamber and the
right guide rail has not changed.
Figure 7-1. Align Frame to Right Guide Rail
Door and Chamber Subsystem
Calibration
7-8
STERRAD® 100 Sterilizer
ASP Service Guide rev D