10 Attach the bellows between the throttle valve and the elbow. Adjust
the pump, or elbow, for optimum alignment. There should be
minimal flexing of the bellows. Us the centering rings and clamps to
secure.
Replace the Bellows
10. Attach the bellows between the throttle valve and the elbow. Adjust
the pump, or elbow for optimum alignment. There should be
minimal flexing for the bellows. Use the centering rings and clamps
to secure.
Replace the Vacuum Switch
11. Attach the vacuum switch using the VCO connector.
Replace the Vent Valve
11. Attach the vent valve assembly.
12. Orient the assembly with the solenoid opposite the HI VAC side of the
throttle valve and parallel to the bellows. Secure the VCO fitting.
Replace the Exhaust Filter
13. Attach the vacuum pump exhaust filter to the pump exhaust. The
drain plug should be on the side clear of the pump and the flow
arrow should point up.
Replace the Oil Return Valve
14. Attach the oil return valve to the pump using the VCO connector.
Attaching the Power Cables
15. Secure the nut plate straps.
16. Attach the harness adapter to the plug end of the vacuum pump
cable. Plug the connectors together and join the ground lugs with the
external tooth lockwasher trapped between them. Secure with the
flat washer and lock nut. The proper connectors are included in the
conversion kit.
Vacuum Subsystem
Removal/Replacement Procedures
4-14
STERRAD® 100 Sterilizer
ASP Service Guide rev. D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......