Action
Press
START.
Display
VACSW AT ATM. SET P1-P4=0
ON SMC EXIT W/ CANCEL
Action
4
4
Note:
If the SMC switch is set at any value other than P1=0,
P2=0, P3=0, P4=0, the machine sounds a beep tone.
F
Use a fine pointed tool to press the vacuum switch SET button. The
LCD displays P1=XX. Use the UP and DOWN keys to set P1=0. ·
F
Press the SET button again and the LCD displays P2. Set P2=0
Repeat to set P3=0 and P4=0. Press SET. The red and green LEDs
turn ON.
Display
SET P1&P3=70, P2&P4=0
ON SMC EXIT W/ CANCEL
Action
4
4
Note:
Ignore the above reading.
F
Use a fine pointed object to press the vacuum switch SET button.
The LCD displays P1=XX. Use the UP and DOWN keys to set P1=70.
F
Press the SET button again and the value of P2 is displayed. Set
P2=15. Repeat the above procedure to set P3=70 and P4=15. When
correctly reset, the beeps stop and the red and green LEDs turn OFF.
Corrective Action
F
Verify that the pneumatic is at the current revision and the
connector jumper on the board is at HI AT ATM. (If the board does
not have a part number, or is not the current revision, this step does
not apply.)·
F
Press RESET on the vacuum switch for 2 seconds.
F
Replace the vacuum switch.
F
Replace the wiring.
Sensor Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-13
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......