RF amplifier failure. Call ASP Technical
Service.
Problem/Condition
F
RF delivered power is less than 100 W.
4
4
Note
: This message does not appear on the STERRAD 100 S
Sterilizer.
Action
F
Check the load and make sure it is not in contact with the electrode.
F
Check the shelves and make sure they are not in contact with the
door, or the chamber.
F
Check the RF power and replace if necessary.
F
Replace the RF amplifier
F
Replace the RF match.
F
Replace the electrode
F
Replace the microprocessor board.
START OPEN DOOR TIME-OUT. CALL
TECHNICAL SERVICE.
There is no printed message for this condition.
Problem/Condition
More than 15 seconds passed and the top sensor has not cleared.
Action
Check the door top sensor.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-33