4. Remove the “Remove Before Use” battery insulator between the
battery and contact of the microprocessor board and affix it to the
DHR.
5. Carefully attach the microprocessor board inside the to control
enclosure.
6. Ensure that all boards are firmly seated in the card cage. Make sure
the safety switch cannot be activated.
System Reset
1. Turn the circuit breakers ON.
2. Press and hold
CANCEL
and at the same time close the control
enclosure cover to activate the safety switch and turn the unit on.
3. Continue to hold
CANCEL
until the machine beeps.
4. Fully engage the enclosure cover screws and lock finger tight.
Verify Voltages
1. Measure AC voltages for chamber and door heaters, control
enclosure, RF generator, vaporizer and compressor. Refer to chart
for version main line voltage and nominal voltages expected ±10% for
each circuit.
International System Configurations
System Installation
ASP Service Guide rev. D
STERRAD® 100 Sterilizer
11-19