CYCLE 2 LONG CYCLE
YES(START)—NO(CANCEL)
F
If you are processing a flexible scope, press
START
to start a long
cycle.
F
If you are not processing a flexible scope, press
CANCEL
to go to the
CYCLE 1 (
short cycle option.
Door Going Up and Down
Problem/Condition
Equipment has been serviced and now door doesn’t work.
Action
F
Check connector on top of optics board. It may have been knocked
loose during servicing.
F
Check for foreign objects under the counterweight.
F
Check door sensors.
F
Perform door test.
Door Stuck in Closed Position
Problem/Condition
Door sensors pass tests but door cannot be opened by either throwing
the circuit breaker or using AutoTest. Door solenoid light is lit (red).
Pressure gauge is in range (60-80 PSI) and compressor is operational.
There is NO vacuum in the chamber.
Action
F
Take the pneumatic door cylinder pressure hoses and switch them.
(One hose blows air.) Door should open but cannot now be closed.
F
Replace door pneumatics solenoid.
F
Check vacuum switch.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-5
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......