Error Handling
If the cassette is inserted incorrectly, or if the cassette sensor finds no
bar code within the time allotted, the cassette is ejected to the operator
with the appropriate message displayed on the LCD. If the bar-code is
located, but indicates a cassette out of date or incorrect cassette type,
the bar-code is read a second time to verify the data.
If the bar-code is read twice as an incorrect type, the cassette is ejected
to the operator. If the bar-code is read twice as cassette out of date, the
cassette is passed to the cassette collector box.
Power Failure
If the power fails during injection, or the first 4 minutes of diffusion, the
cassette is positioned at the next available cell when the power is
restored. If the cassette is at a mark, it does not advance until after the
next attempted injection.
If power fails when the system is idle, the cassette is advanced to the
next position when power is restored.
Injection Subsystem Calibration
IMPORTANT!:
Your first step when troubleshooting a system is to verify
that the system is
properly connected to the power source
, and to
make sure all the
subsystems are properly connected to the AC
system and all circuit breakers are on and functioning normally. Once
you have verified these things, proceed with your troubleshooting steps.
The injection subsystem, consisting of (for the purposes of this guide)
the injector pump, chute and cassette collection box should be
calibrated when the unit is replaced, when there is a malfunction as
detailed in the removal procedures, when product certification shows
faults in the system.
WARNING! H
2
O
2
EXPOSURE! Before touching the injector
valve, be sure there is no sign of H
2
O
2
residue. If
there is, be sure to wear gloves to prevent H
2
O
2
exposure.
Injector Valve Height
1. Manually extend the injector cam and insert the cassette gap tool
(injector valve gage) from the back of the platen. Position it on the
injector valve seal retainer plate over the injector valve needle seals.
Injection Subsystem
Injection Subsystem Calibration
5-4
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......