Display
PRESS= XXX, FWD= XXX
RFL= XXX, OFF W/ START
Action
F
Allow pressure to stabilize, observe pressure and RF forward voltage
reading on control panel LCD display. ·
F
Verify that pressure is stable every 5 minutes for 15 minutes. Note
RF power reading
F
RF delivered (from LCD read-out) - 3000 mVDC minimum.
F
RF reflective is less than 300 mVDC.
F
RF power forward from power meter - .360 kW, .440 kW.
F
After running plasma for 15 minutes, press START and verify the RF
generator is off by verifying that the Watt meter reads 0.·
F
Turn off the Watt meter and disconnect it from the RF generator.
F
Remove the RF interconnect cable from the power meter and
reconnect it to RF generator.
F
Press
START
to verify voltages are correctly displayed and to verify
that plasma is ON.
F
Press a
LONG
CANCEL
from this menu to perform a leak test.
Corrective Action
F
Check the connections at the RF match, cable and electrode.
F
Check the electrode screws and match screws are correctly seated.
F
Replace the electrode if necessary.
F
Replace the RF delivery system if plasma is not induced or RF power
failure.
Leak Back Test
Display
LEAKTEST W/ START
NEXT W/ LONG CANCEL
Action
Press
START
once.
Vacuum Subsystem Tests
AutoTest
ASP Service Guide rev D
STERRAD® 100 Sterilizer
B-37
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......