Chamber/Door Heaters Test
Display
ON/OFF HEATER
W/ START OR NEXT W/ CANCEL
Action
Press
START.
Display
HEAT ON W/ START
NEXT OUTPUT W/ CANCEL
Action
F
Press
START.
F
Check that voltage at chamber heater connectors (1 chamber
connector on Block 2) reads 115 or 230 VAC depending on system
type. (208 VAC for Block 2.)
F
Check that voltage at door heater connectors P95 reads 115 or 200
VAC depending on system type. (208 VAC for Block 2.) These are
nominal readings only to check shorts and opens. (There is only one
connector on Block 2.)
Corrective Action
F
Replace thermistors.
F
Replace solid state relay board.
F
Verify impedance of heaters.
4
4
Note:
If temperature error is shown on the display, press
OPEN to recover. Repeat the test.
Display
HEAT OFF W/ START AXXX.XV
NEXT OUTPUT W/ CANCEL
Action
F
Press
START
.
F
Verify voltage at P115 in the chamber reads 1 VAC. There is only
one connector on Block 2.)·
AutoTest
Throttle Valve/Heater Menus
B-24
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......