Remove/Replace the Firmware
CAUTION! STATIC SENSITIVE DEVICES. You must use a static
strap and have a static controlled area available
when performing this procedure.
1. Examine the last cycle printout. Note the total machine cycles.
2. Press
START
and
CANCEL
at the same time. The system displays the
current date.
3. Press
CANCEL
twice. Verify that the total machine cycles (next to the
NEXT PM field) are the same as the printout.
4. Turn main power circuit breaker OFF.
5. Remove the control enclosure panel.
6. Remove the existing microprocessor board and place it on a static
controlled work area. Remove the exiting EPROM from the board and
install the replacement firmware. Pay close attention to the
orientation of the chip.
7. Replace the microprocessor board.
8. Turn the main power circuit breaker to ON.
9. Perform a Master Reset.
Figure 6-7. Power Supply Wiring
User Interface and Control Subsystems
Control Subsystem Replacement
6-26
STERRAD® 100 Sterilizer
ASP Service Guide rev D.