There is no printed message for this condition.
Problem/Condition
F
The inserted cassette is not the correct type for the STERRAD® 100
Sterilizer.
F
The inserted cassette is not the correct type for the STERRAD
100 S
Sterilizer.
Action
F
Insert a STERRAD 100 Cassette.
F
Insert a STERRAD
100 S
Cassette
INDEX CASSETTE FAILURE CALL
TECHNICAL SERVICE
There is no printed message for this condition.
Problem/Condition
The system failed to detect the cassette. The cassette is ejected into the
collector box.
Action
F
Inspect the cassette, the pump optics and electronics.
F
Perform cassette sensor test.
F
Remove and replace injector pump.
F
Check the floating guides