POSITIONING CASSETTE
There is no printed message for this condition.
Problem/Condition
The cassette is being positioned. The reflective sensor did not detect the
side cell positioning mark. The cassette advances to the next position.
Action
F
Inspect the cassette for damage or smudges on bar-code.
F
Insert a new cassette.
F
Test the pump optics and electronics
Power interrupted.
Problem/Condition
Possible brief power failure.
Action
Check customer log to determine if there is an ongoing problem in this
area. If this is a chronic problem, site may have chronic power problems
that need to be addressed by ASP Technical Services and customer’s
engineering department.
Power interrupted. Call ASP Technical
Service.
Problem/Condition
Power was interrupted for the second time and for more than 1 minute.
Action
F
Check incoming power supply. Verify that phase is correct.
F
Check for loose connections.
F
Replace the AC enclosure
System Messages
Using System Messages
3-24
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......