Injection system interrupted. Repackage
load. Restart sterilizer. Insert new
cassette.
Problem/Condition
The sterilization process canceled because the cassette was faulty.
Action
The faulty cassette ejects into the collection box.
F
Insert a new cassette.
F
Repackage the load.
F
Restart the sterilizer.
Injector heater failure. Call ASP
Technical Service.
Problem/Condition
The current sensor has not detected activation of the vaporizer heater.
Action
F
Check the vaporizer heater.
F
Inspect the injector valve and replace if necessary.
F
Replace the current sensor.
System Messages
Using System Messages
3-8
STERRAD® 100 Sterilizer
ASP Service Guide rev D