G
Injector valve
G
Quartz window
G
Bellows and Vacuum O-ring clamps
G
Door seal O-ring
G
Vent valve
G
Throttle valve
G
RF feedthrough
Continuous Positive BIs
Problem/Condition
The biological Indicator test is repeatedly positive. The sterilizer is
running normally in all other respects and the operator is performing the
tests correctly.
Action
F
Ensure that vaporizer cap is free from damage and does not show
signs of abnormal wear. Clean and or replace as necessary.
F
Check injector valve height.
F
Check cassette associated with the suspect process cycle. Open
cassette and inspect for damage, residual H2O2, or physical
abnormality.
F
Review process parameter printer tape for suspect process cycle.
Verify recorded injection pressure is correct.
F
Review cycle history for previous 30 days. Report any process
parameter printouts and records for any previous positive or suspect
positive cycles.
F
Make sure electrode screw is tight.
F
Make sure all RF system connectors and screws are tight.
F
Replace the electrode.
System Messages
Using System Messages
3-44
STERRAD® 100 Sterilizer
ASP Service Guide rev D
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......