OVER CENTER TIME-OUT CALL TECHNICAL
SERVICE.
There is no printed message for this condition.
Problem/Condition
The door reached the top sensor but did not trip the over-center sensor
within 15 seconds.
Action
F
Check the over center sensor.
F
Replace the over-center sensor.
F
Check injector pump harnesses.
Over pressure in diffusion. Call ASP
Technical Service
Problem/Condition
F
For the second consecutive cycle, diffusion pressure is over 14 Torr,
or;
F
During empty chamber cycle a steady rise in chamber pressure is
seen in injection stage and then in diffusion stage with sudden
shut-down in diffusion. Injection pressure is in range, vacuum stage
pressure is good, divide by 2 voltage is good.
Action
F
Run an empty chamber cycle.
F
Examine the chamber and the door seal for any possible leaks.
F
Perform a leak back test.
F
Replace the analog board if the second problem is seen. This is NOT
a leak.
F
Replace the capacitance manometer.
Using System Messages
System Messages
ASP Service Guide rev D
STERRAD® 100 Sterilizer
3-19
Содержание STERRAD 100
Страница 284: ......
Страница 285: ......
Страница 286: ......
Страница 287: ......
Страница 288: ......
Страница 289: ......
Страница 290: ......
Страница 291: ......
Страница 292: ......
Страница 293: ......
Страница 294: ......
Страница 295: ......
Страница 296: ......
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......
Страница 304: ......
Страница 305: ......
Страница 306: ......
Страница 307: ......
Страница 308: ......
Страница 309: ......