VENT TIME-OUT CALL TECHNICAL SERVICE
Vent time-out. Call ASP Technical Service.
Problem/Condition
Atmospheric pressure was not detected by the end of the vent stage in
vacuum, diffusion or at the end of the sterilization cycle.
Action
F
Check time on Whitman Vacuum Switch and reset if necessary.
F
Check LCD during vacuum switch test Verify that setting is O. If
not, reset for 2 seconds.
F
Make sure vacuum switch retaining collar is tight.
F
Make sure vacuum switch is operational.
F
Check DC voltages
F
Replace vacuum switch
F
Replace pneumatic board
VERIFYING CASSETTE PLEASE WAIT...
There is no printed message for this condition.
Problem/Condition
The system is trying to read the bar code.
Action
F
Insert a new cassette.
F
If the message persists, test the cassette optics.
System Messages
Using System Messages
3-42
STERRAD® 100 Sterilizer
ASP Service Guide rev D