Action
F
Press
START
. Verify both plasma lamps are on.
F
Press
START
to turn light off.
F
Press
CANCEL
for light test.
Display
ON/OFF LIGHTS W/ START
NEXT LIGHT W/ CANCEL
Action
F
Press
START
. Verify both vent lamps are on.
F
Press
START
to turn light off.
F
Press
CANCEL
for next screen test.
Corrective Action
F
Check bulbs and replace if necessary.
F
Check control signal from CPU to control panel board.
F
Check/replace control panel assembly.
Sensor Tests
Display
SENSOR TEST W/ START
NEXT Menu W/ LONG CANCEL
Action
F
Press
START
.
AutoTest
Sensor Tests
B-8
STERRAD® 100 Sterilizer
ASP Service Guide rev D