Firmware Screen Descriptions
View Records Menu
9-16
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Time 2 of Max Temp (hh:mm)*
PM 10 Sample Volume*
Power Failure Date (yyyy/mm/dd)
Power Failure Time (hh:mm)
*
2000
i
-D only
The Sampler stores the following as 5-minute interval data.
Ambient Temp
Filter Temp
Ambient Pres
Flow
Filter2 Temp*
Flow2*
*
2000
i
-D only
It writes a new record of interval data every 5 minutes on a continuous
basis, and has months of capacity before it overwrites the oldest records.
●
In the Main Menu, choose View Records >
Interval/Srec
.
time date flags
04:40 05/04/10 20130100
04:45 05/04/10 20130100
04:50 05/04/10 20130100
04:55 05/04/10 20130100
PGUP/DN
PAN L/R
CTRST SSET INSTS STAT
The User/Lrec screen displays averaged meteorological and input data at
the interval specified by the user. The default averaging/storage interval is
one minute. The Sampler has the capacity to retain more than 17 days of
input data (at one minute logging of 18 data items) before it overwrites the
oldest records.
Default setup:
Flow
Flow2*
Ambient Pres
Filter Pres
Filter2 Pres*
Pump Pres
Ambient Temp
Filter Temp
Interval/Srec
User/Lrec