C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
B-29
Table B–8.
Set Layout Ack Values
Value
Function
0
Do nothing (default)
1 Append
“*”
tz
This command reports the “tz” timezone string for the NTP server. See
“Network Time Protocol Server” in the “Communications Settings”
section of the “Operation” chapter for more information.
Send:
tz
Receive: tz
EST+5EDT
set tz
string
This command sets the timezone
string
for the instrument for use with the
NTP server, where
string
is a standard timezone string. Common strings are
listed in the “Timezone” screen description in Chapter 3.
Send:
set tz EST+5EDT
Receive:
set tz EST+5EDT ok
analog iout range
channel
This command reports the analog current output range setting for
channel
,
where
channel
must be between 1 and 6, inclusive. The following example
reports current output channel 4 is in the 4-20 mA range, according to
Table B–9
. This command responds with “feature not enabled” if the I/O
expansion board is not detected.
Send:
analog iout range 4
Receive:
analog iout range 4 2
set analog iout range
channel range
This command sets analog current output
channel
to the
range
where
channel
is between 1 and 6 inclusive, and
range
is set according to
Table B–
9
. The following example sets current output channel 4 to the 0-20 mA
range. This command responds with “feature not enabled” if the I/O
expansion board is not detected.
Send:
set analog iout range 4 1
Receive:
set analog iout range 4 1 ok
I/O Configuration