Firmware Screen Descriptions
Diagnostics
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-91
VOLTAGES:
>MOTHERBOARD
INTERFACE BOARD
I/O BOARD
CTRST SSET INSTS STAT
The Motherboard Voltages screen (read only) is used to display the voltage
readings on the motherboard.
●
In the Main Menu, choose Diagnostics
>
Voltages >
Motherboard
.
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.1 V
15.0 SUPPLY 15.2 V
24.0 SUPPLY 24.1 V
-3.3 SUPPLY -3.3 V
CTRST SSET INSTS STAT
The Interface Board Voltages screen (read only) is used to display the
voltage readings on the measurement interface board.
●
In the Main Menu, choose Diagnostics
>
Voltages
>
Interface Board
.
INTERFACE BOARD VOLTAGES:
> 3.3 SUPPLY 3.4
5.0 SUPPLY 5.1
15.0 SUPPLY 15.2
24.0 SUPPLY 24.2
CTRST SSET INSTS STAT
The I/O Board Voltages screen (read only) is used to display the voltage
readings on the I/O expansion board. This menu is displayed only if the
optional I/O expansion board is installed.
●
In the Main Menu, choose Diagnostics
>
Voltages >
I/O Board
.
Motherboard Voltages
Interface Board Voltages
I/O Board Voltages