Firmware Screen Descriptions
Instrument Setup Menu
9-26
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
The Filter Fan screen controls the filter fan operating mode. If this field is
set to Always On, the filter fan will run continuously. If this field is set to
Automatic, the filter fan will only initiate if the temperature of the filter
and ambient air differs by 2 °C. If the Filter Fan field is set to Always Off,
the filter fan will not initiate at all. When you set this field to Always Off,
the filter temperature range error code (that initiates the filter fan in the
Automatic setting) will be disabled.
●
In the Main Menu, choose Instrument Setup > System Setup >
Filter
Fan
.
FILTER FAN MODE:
CURRENTLY: AUTOMATIC
SET TO: AUTOMATIC
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Auto Run screen is used to set the Auto Run mode to On/Off. If no
keys are pressed for three hours and the Auto Run feature is On, the unit
will automatically enter the Wait or Sampling mode from the Stop mode.
●
In the Main Menu, choose Instrument Setup > System Setup >
Auto
Run
.
AUTO RUN MODE:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
CTRST SSET INSTS STAT
The Flow Calibration Setup menu is used to setup flow calibration. Flow 2
is used with the 2000
i
-D.
●
In the Main Menu, choose Instrument Setup >
Flow Calibration
Setup
.
Filter Fan
Auto Run
Flow Calibration
Setup