AK Protocol
Commands List
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
E-11
AK Protocol
Set Storage Marker Command (SSTO)
COM 2-WAY SETTINGS
RS-Para 1
52
ASCII code for the 1-digit Station Number (for example "4": 052). The Station Number is always 1 digit in length.
RS-Para 2
75048
ASCII code representation of the 2-digit Channel Number (for example: "K0": 075, 048). The Channel Number is
always 2 digits in length.
RS-Para 3
13010
Optional: Up to 3 ASCII codes can be added to response from instrument. In this case, <CR> and <LF> (ASCII
codes 013 and 010) are appended to the response. Enter 0 if nothing is to be appended.
RS-Para 4
0
Not used.
Transmission to Instrument
Response from Instrument
Byte
Example
Description
B
No Err
Error
Description
1
<STX>
ASCII code 002.
1
<STX>
<STX>
ASCII code 002.
2
4
1-digit Station Number, RS-Para 1.
2
4
4
1-digit Station Number, RS-Para 1.
3
S
4-digit Set Storage Marker
command.
3
S
S
4-digit Set Storage Marker command.
4
S
4
S
S
5
T
5
T
T
6
O
6
O
O
7
<space>
Space.
7
<space>
<space>
Space.
8
K
2-digit Channel Number, as
defined by RS-Para 2.
8
0
0
Number of current status conditions.
9
0
9
<ETX>
<space>
ASCII code 003.
10
<space>
Space.
10
<CR>
S
Up to 3 digits appended to the end of
the response transmission, according
to the entry for RS-Para 3.
11
B
New location of the Storage
Marker.
B: move to beginning of storage
buffer, E: move to end of storage
buffer. Enter positive numbers
such as 250 to move forward by n
records, and negative numbers
such as -1000 to move backwards
by n records. Do not right-fill.
11
<LF>
E
12
12
<ETX>
13
13
<CR>
14
14
<LF>
15
15
16
<ETX>
ASCII code 003.
16
17
17
18
18
19
19