Calibration
Flow Calibration
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
8-11
5.
Scroll to Flow to perform the calibration of the 20 LPM or Fine
Sample channel.
6.
Open the filter exchanger and place the Flow Calibration Cassette in
the Coarse (Front) position of the dual holder. Place a regular cassette
with filter media in the Fine (Back) position of the dual holder.
Figure 8–2.
Flow Calibration Cassette in the Front Position
7.
Place the dual filter holder in the filter exchange mechanism and close
the filter exchange mechanism making sure the seals are not
compromised.
Figure 8–3.
Dual Filter Holder in the Filter Exchange
Figure 8–4.
Filter Exchange Closed