Preventive Maintenance
V-Seals Cleaning and Replacement
10-16
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Figure 10–10.
Pulling Backward on the Handle of the Filter Exchange
Mechanism
4.
Release the handle and allow it to drop down.
5.
Remove the filter cassette carrier from the filter platform. If the sample
has been interrupted for an audit, store the filter in the cassette carrier
box until ready to resume sampling.
6.
Push the filter exchange mechanism toward the back slightly, and lift
the rollers through the slots in the left-hand and right-hand guides
(
Figure 10–11
).
Figure 10–11.
Lifting the Rollers through the Guide Slots
7.
Allow the filter platform to settle downward.