Servicing
Firmware Updates
12-2
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
touching any internal components. When the instrument is unplugged, the
chassis is not at earth ground.
▲
Do not use solvents or other cleaning products to clean the outside case.
▲
Do not remove the LCD panel or frame from the LCD module.
▲
The LCD polarizing plate is very fragile, handle it carefully.
▲
Do not wipe the LCD polarizing plate with a dry cloth, as it may easily
scratch the plate.
▲
Do not use alcohol, acetone, MEK or other Ketone based or aromatic
solvents to clean the LCD module, but rather use a soft cloth moistened
with a naphtha cleaning solvent.
▲
Do not place the LCD module near organic solvents or corrosive gases.
▲
Do not shake or jolt the LCD module.
▲
The firmware can be updated by the user in the field via the serial port,
over the Ethernet, or from a file on a USB memory device. This includes
both the main processor firmware and the firmware in all low-level
processors. Refer to the
iPort
manual for the firmware update procedure
when using the serial port or Ethernet. To update the firmware using USB
memory device, use the following instructions:
1.
Obtain the firmware update file and copy to the root directory of the
USB memory device. The firmware can be downloaded from the
customer area of the Thermo Fisher Scientific wbesite and should be
copied to the root directory of the USB memory device.
2.
The Sampler must be in Service mode to upload new firmware. Press
, scroll to Service mode and press
to toggle the Service
mode On.
3.
Insert the USB memory device with the firmware update into the USB
port on the Sampler.
Firmware
Updates