Firmware Screen Descriptions
Instrument Setup Menu
9-28
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
SET NUMBER OF POINTS:
CURRENTLY: 5
SET TO: 5
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Reset Flow Cal menu item is used to reset the flow calibration
parameters.
●
In the Main Menu, choose Instrument Setup >
Flow Calibration Setup
> Flow
>
Reset Flow Cal > press
to reset the flow calibration
parameters.
The Flow Constant M/B screen allows the user to adjust the setpoint. The
Flow Constant B screen functions in the same manner as the Flow
Constant M screen shown in the example.
●
In the Main Menu, choose Instrument Setup >
Flow Calibration Setup
> Flow
>
FTS Constant M
.
FTS CONSTANT M:
CURRENTLY: 0.0000
SET TO: 0.000
0
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The
i
Series instruments include a built-in data logging capability as a
standard feature. The operator is allowed to create two different types of
records, which for historical reasons are named lrecs (long records) and
srecs (short records). Each record can contain up to 32 different fields or
data items, and records can be created at user-defined intervals ranging
from 1 to 60 minutes.
Record generation is tied to the instrument’s real-time clock. For example,
if the logging period for srecs is set to 30 minutes, a new srec will be
generated on every hour and every half hour (10:00, 10:30, 11:00 …).
Lrecs and srecs can be interleaved.
The Sampler’s computer system includes three megabytes of flash memory,
which is enough to store a full lrec containing 32 data items and a full srec
containing 32 items once each minute for a week (>20,000 total records).
Reset Flow Cal
FTS Constant M/B
Datalogging