Firmware Screen Descriptions
Service Menu
9-82
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
selecting the range, setting minimum/maximum values, and choosing the
signal to output.
●
In the Main Menu, choose Service > External I/O > Analog Output
Configuration >
All Voltage Channels, All Current Channels,
Voltage Channel 1–6 or Current Channel 1–6
.
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
CTRST SSET INSTS STAT
The Select Range screen is used to select the hardware range for the selected
analog voltage output channel. Possible ranges for the voltage outputs are:
0-100 mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current outputs
are: 0-20 mA and 4-20 mA.
●
In the Main Menu, choose Service > External I/O > Analog Output
Configuration >
select channel
>
Select Range
.
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-100mV ?
CHANGE VALUE
SAVE
CTRST SSET INSTS STAT
The Minimum Value screen is used to edit the zero (0) to full-scale (100)
value in percentages for the selected analog output channel. See
Table 9–9
for a list of choices. The Minimum and Maximum Value screens function
the same way.
●
In the Main Menu, choose Service > External I/O > Analog Output
Configuration > >
select channel
> >
Set Minimum Value
or
Set
Maximum Value
.
Select Range
Minimum and Maximum
Value