Firmware Screen Descriptions
Service Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-87
The Number of Table Points screen allows the user to select how many
points are used in the analog input conversion table for the selected
channel. The instrument uses linear interpolation between the points in
this table to determine what the reading value is, based on the analog input
voltage. Each point in the table consists of an analog input voltage value (0-
10 V) and a corresponding reading value. Only two points are necessary for
linear inputs, however, a larger number of points may be used to
approximate non-linear inputs. The points range from 2 to 10, with a
default of 2.
●
In the Main Menu, choose Service > External I/O > Analog Input
Configuration >
select channel (1-8)
>
Table Points
.
NUMBER OF TABLE POINTS:
CURRENTLY: 2
SET TO: 10 ?
INC/DEC
SAVE VALUE
CTRST SSET INSTS STAT
The Table Point submenu allows the user to set up an individual table
point.
●
In the Main Menu, choose Service > External I/O > Analog Input
Configuration >
select channel (1-8)
>
Point (1-2)
.
TABLE POINT 01 CONFIG:
>VOLTS 0.00
USER VALUE 0.00
CTRST SSET INSTS STAT
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10. The default table is a
two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U, where U is the previously entered unit of measure.
●
In the Main Menu, choose Service > External I/O > Analog Input
Configuration >
select channel
>
select point
>
Volts
.
Number of Table Points
Table Point
Volts