C-Link Protocol Commands
Communications Configuration
B-26
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Send:
instr
name
Receive: instr
name
Partisol
Sampler
Partisol
Sampler
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The following example shows that the instrument is in
the remote mode.
Send:
mode
Receive: mode
remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The
following example sets the instrument to the local mode.
Send:
set
mode
local
Receive:
set mode local ok
allow mode cmd
This command reports the current allow mode setting: 1 = allow “set mode
local” and “set mode remote” commands; 0 = ignore “set mode local” or
“set mode remote”commands, according to
Table B–6
. The default value is
0; ignore the commands. The following example shows that the instrument
is configured to ignore “set mode local” or “set mode remote” commands.
Send:
allow
mode
cmd
Receive:
allow mode cmd 0
set allow mode cmd
value
This command is used to configure the instrument to
value
, where
value
is
either 1 = accept or 0 = ignore the “set mode local” and “set mode remote”
commands, according to
Table B–6
.
If the instrument is set to accept the commands (
value
= 1), the “set mode
local” command will unlock the instrument and the keypad can be used to
make changes via the front panel.
If the instrument is set to ignore the commands (
value
= 0), the instrument
will respond with “ok” as if the command has been accepted and acted
upon,
but it will not change the instrument lock status
(this is for
compatibility with systems expecting and “ok” response).